Dissolvi Ozone Analyser deprehendatur principle: UV Dual iter effusio modum, propter analysis et deprehendatur dissolvi ozone aquam concentration in ICTUS aqua, pura aqua, aut semiconductor industria.
Deprehensio principle: UV Dual Semita effusio modum, propter analysis et deprehendatur dissolvi ozone aquam concentration in ICTUS aqua, pura aqua, aut semiconductor industria.
Mensurement range: 0-20PPM; 0-50PPM; 0-100ppm; 0-120PPM; 0-10DPM; 0-200ppm.
Product Features: Hoc UV Dissolvi Ozone Analyser habet internum lucem fons automatic disparet munus, quod potest adjust claritas ducitur lux fons in real-tempus fundatur in lucem notitia de nulla puncto ad curare de stabilitatem in nulla parte. UV optical ozone aquam concentration analyser potest connexa in parallela aut serie ad dissolvi ozone aquam Pipeline, maxime propter analysis et deprehendatur dissolvi ozone aqua concentration in ICTUS aquae, pura aqua vel semiconductor in ICTUS aquae, pura aqua vel semiconductor.
Haec dissolvi ozone analyser habet automatic temperatio ratio munus cum UV lux source, quod potest adjust in claritas in UV ducitur lux fons in real-vicis secundum referat lucem data est ponere constans nulla puncto ad secundum rationem ponere constans nulla punctum, ensuring in stabilitatem in nulla parte. Est patented uber nostrae company. Hoc UV optical OZONE Aqua incumbo Analyser potest operari continuously sine prohibitione pro XXIV horas.
Dissolvi Ozone concentration Analyser est a recta mensura OZONE contentus in ultrapure aqua vel constant TEGENS AQUA UTRANVIOLET spectrophotometry. Et customizable version est HF, quod potest adhiberi ad XX% hydrofluoric acidum ozone aqueum solutio. Materials in contactu cum OZONE aqua includit PTFE, FFPM, PFA et Vicus vel Sapphire. In UVEZ, 180w dissolvi OZONE Aqua incumbo analyser est instructa cum in lente et exitus, et instructa tactus screen operationem interface. In output est directe coniuncta ad PLC cum 4-20MA RS-CDLXXXV signum lineas.
Et dissolvi ozone concentration analyser habet automatic temperatio ratio munus cum UV lux source, quod potest adjust claritas in UV ducitur lux source in real-vicis secundum referat lucem data est ponere constans nulla puncto ad lucem data est ponere constans nulla punctum. Est patented uber nostrae company. Hoc ozone analyser potest operari continuously sine prohibitione pro XXIV horas. Superior stabilitatem huius ozone concentration analyser est effectum per usum verus dual trabem optical system et maxime diu-vixerunt ultraviolet lux fonte.
Crusta dissolvi ozone concentration analyser est aluminium Crusta cum RESPERGO probationem consilio, praesidio gradu: iP65, size: 260x160x90mm (WXHXD) et pondus de III kilograms. Haec dissolvi OZONE Aqua concentration Analyser potest esse murum mounted cum hamo consilio tergo. Cum installation non requiritur, quattuor ascendens uncis sunt provisum.
In livore fluxus rate de dissolvi ozone aquam concentration analyser est 200-300ml / min. Si fluxus imperium non requiritur, a suffocare valvae vel fluxus meter est installed ad exitus de dissolvi ozone aquam concentration analysar. This IUGULUS CYMBALE sit positus post sensorem (nunquam ante eam? Installing a suffocare valvae ad incole generate bullae, ita stricte prohibetur ozone aqua analyzer inire dissolvi aquam.)
Oportet ponere munditia aquae dissolvi OZONE Aqua incumbo Analyzer et tenere munditiam internis sensoriis dissolvi ozone aqua analyzer. Est suadetur ut calibrate in nulla parte dissolvi ozone aqua concentration analyser omnibus II weeks. Cum calibrating nulla puncto, mundus sine ozone vel color debet introduci
Test modum: Dual Semita UV effusio modum, diu-vita lux source ratio, excelsum mensurae accurate.
Mensurae principle: Secundum Lambertus Cervus lege, precise mensurae peragitur per principium photometric effusio.
Mensurement range: 0-20PPM; 0-50PPM; 0-100ppm; 0-120PPM; 0-10DPM; 0-200ppm.
Lux Source System: Aliena Long-Vita Ultraviolet lux Source System (253.7nm), cum libera warantum in III annis.
Usus: UMBRILIUM installation et usu instructa cum IMPERVIUS Aviation obturaculum terminals.
Lux stagnum ratio: separatum lux stagnum technology, Leak liberum, princeps pressura resistentia, et resistentia ad altum fluunt rate sampling liquido impulsum.
Intelligentes Compensation: instructa cum ipso nulla temperatio et Lorem lux Source ultricies munus.
Operatio modum: quod potest esse manually zeroed, et durante zeroing, necesse est ut non sit ozone aqueum solutio in instrumenti.
Display Unit: MG / L, PPM libitum.
Data Display: High Definitio color tactus screen.
Output munus: 4-20ma, Rs485 Communication.
Automatic nulla temperatio: Non potest adjust claritas ad ultraviolet DUXERIT lux Source in Real-Time fundatur in referat lucem notitia ponere constans nulla punctum data, cursus stabilitatem in nulla parte.
Multiple Versions:
In normalis version adhibetur ad ICTUS aqua et purificatus aqua.
Et HF version potest esse pro XX% hydrofluoric acidum ozone aqueum solution.
【I】 Wall mounted installation dimensiones
【II】 Display Panel et Display collaborative
【III】 Power input, signum output, in litt
Power initus:
(I) 24V DC + |
(II) 24v DC- |
Signum Terminal output:
(I) Rs485 + (Modbus) |
(IV) 4-20ma- output |
(II) Rs485- (Modbus) |
(V) zeroing + initus |
(III) 4-20ma + output |
(VI) zeroing- input |
Dissplaplay Signum Terminal: solebat coniungere Longinquus display annuit (haec apparatus iam habet ostentationem screen)
In Gas interface super UV dissolvi ozone aqua analyzer adhibetur ad compressum aerem interface. Si temperatus OZONE aqua est humilis, commendatur ad introducendam compressum aer ad instrumentum ne densam in internum ozone aqua Pipeline instrumenti.
Dissolvi Ozone Analyser - positivum pressura genere sampling nexu diagram
(Est commendatur ut hoc nexu processus pro split sampling. Nota quod fluxus meter necessitates ad installed in exitus de dissolvi ozone aqua analyzer.)
Dissolvi Ozone Analyser, negans aquas sentinam sampling connexionem diagram
(Est commendatur ut hoc nexu processus pro split sampling. Nota quod fluxus meter necessitates ad installed in exitus de dissolvi ozone aqua analyzer.)
Nota: De manufacturer de dissolvi ozone aqua fluxus meter est a vexillum configuratione accessorium et negative pressura aqua sentinam est ad libitum uber. Vos can consule fabrica ut emo quod seorsum.
Nota: Sine Display version, in manufacturer et separatim configurare wired tactus screen accessiones (filum longitudinem in II metris)